Downloaded from https://www.everyspec.com MIL-STD-1311D 5.3 Precautions. 5.3.1 A level signal source is recommended where possible. 5.3.2 The calibration line will not be linear with VSWR. It is desirable to use calibrated mismatches of the same order of magnitude as the mismatch of the TUT. 5.3.3 Errors may be introduced because of inadequate directivity in the directional couplers. This is particularly important in measuring low VSWR's. The directivity of the coupler shall be at least 20 dB better than the return loss from the TUT to avoid errors due to phase effects on the mismatch. 5.3.4 The signal source shall be sufficiently decoupled from its load.

5.3.5 Hot input and output matches may depend upon signal level because of tube non-linearity. 6. Method 2B (comparison to calibration short circuit). Refer to figure 4256-2 in which a calibrated variable attenuator is inserted in the directional coupler leg at B at a fixed output from the signal source. With the TUT replaced by the calibration short, a baseline and calibration lines are established on the face of the CRO or recorder. The VSWR calibration lines are related to the attenuator setting by the following formula (if the measurement to be taken is a power measurement): Test procedure and precautions remain the same as 5.2 and 5.3. 7.

Method 3 (manual hot output match with applied input signal). 7.1 Test circuit. 1 1 log 10 + = VSWR VSWR dB in setting Attenuator FIGURE 4256-3. Manual hot output match. METHOD 4256 4256-3 19 April 1968