Downloaded from https://www.everyspec.com MIL-STD-1311D FIGURE 4277-2. Typical plot of power versus frequency. 2.3 The average power, P, within the specified bandwidth is determined by the relationship, At Po x A P = Where: A = Total area within bandwidth specified. At = Total area under plot. Po = Total average power output from TUT. 2.4 The bandwidth (BW) at a specified power density level (pdl) is obtained directly from the calibrated recorder plot as illustrated on figure 4277-2. Calibrating in this fashion the power meter reading corresponding to Y can be used without the recorder to determine the two frequencies at which Y is first reached. 2.5 Precautions. a.

The tunable filter shall have a 3 dB bandwidth that is less than 10 percent of the bandwidth being examined. b. The directional coupler shall have a relatively flat response or known calibration over the frequency range of interest and harmonics should be excluded from the thermistor. 3. Method B. Average tuning sensitivity/spot tuning sensitivity method. 3.1 In method B, power density is defined as the power per unit bandwidth delivered when the tube is tuned across the specified frequency band by a linear voltage sweep. It is, therefore, affected by both power output and tuning sensitivity variations across the band, which are allowed for in the following method.