Downloaded from https://www.everyspec.com MIL-STD-1311D FREQUENCY PUSHING FIGURE 1. Purpose. This test is performed to determine the change in frequency of oscillation resulting from a change in electrode current. 2. Equipment. a. Power sources required for operation of the tube and a specified matched load capable of absorbing the output power. b. Either spectrum analyzer or wavemeter to measure the oscillation frequencies. c. Modulation equipment capable of varying the current over the specified range, either sinusoidally or by switching between two levels at a frequency not less than 45 Hz or as specified. 3. Procedure.

The frequency pushing figure shall be measured by varying the electrode current periodically about the operating point under the specified operating conditions. The difference between the extremes of frequency of oscillation throughout the specified current variation shall be measured and the frequency pushing figure computed as the ratio of (1) this difference in frequency, to (2) the magnitude of the current variation. One of the following methods shall be used as specified. 3.1 Method A (pulse operation). a.

The amplitude of the electrode current pulse shall be periodically modulated, usually sinusoidally, at such a rate that thermal effects may be neglected, with provision being made to measure the peak difference in the changing current. The frequency change may be measured by a spectrum analyzer or by a receiver. b. The result is to be expressed as the ratio of (1) the change in frequency, to (2) the difference between current peaks during a modulation cycle. 3.2 Method B (CW operation). a. The electrode current shall be modulated smoothly over, at least the specified range, at such a rate that thermal effects may be neglected. b.

The output of a suitable frequency discriminator shall be connected to one axis of the oscilloscope and the electrode current signal applied to the other axis which shall be calibrated in order that the instantaneous frequency at all values of current can be observed. 4. Precaution. The frequency discriminator shall be of sufficiently high resolution to resolve the expected frequency difference. METHOD 4311A METHOD 4311A 4311-1 25 March 1970