Downloaded from https://www.everyspec.com MIL-STD-1311D Where: Fo = Resonant frequency. df = F1-F2. b = B/Yo = Normalized susceptance of the tube. g = G/Yo = Normalized conductance of the tube. Since Fo is always close to Fr, (the specified reference frequency for the tube), this latter quantity shall always be used in place of Fo in the above equation. The quantity dB , which is the rate of change of susceptance with frequency, in the vicinity of Fo, can be df determined by the measurement of b at two frequencies near Fo and assuming a linear relationship between b and f. The linear relationship may be assumed for any frequencies within 1 percent of Fo.
An alternative technique for the determination of df dB involves the measurement in front of the tube of the rate of change of the phase of the standing wave minimum with frequency, then: df d g g df dB 1 ) 2 1( 2 + = λ π Where: λ g = Guide wavelength at the reference frequency. g = Normalized conductance of the tube. df d 1 = Rate of change of the phase of the voltage standing wave before the tube with respect to frequency. df d 1 shall be determined at the voltage maximum close to the plane of symmetry of the tube. Since it is not usually possible to take measurements at this position, impedance shall be measured at a remote point and a correction for the length of line shall be made.
This is: g df d m df d df d λ 4 '1 1 = Where: df d '1 = Measured slope of the line obtained by plotting the observed position of a voltage minimum as a function of frequency. m = Odd number of quarter wavelengths measured, at resonance, from the reference minimum to the plane of symmetry of the tube. Frequencies within 1 percent of Fo should be used in the measurement of df d 1 . METHOD 4461A 4461-2 25 March 1970