Department of Defense MIL-STD-1311C — MIL-STD-1311C Test Methods for Electron Tubes – DoD Standard PDF

Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 METHOD 4250A POWER LEVELS (Microwave tubes) 1. Purpose. The purpose of these tests is to determine power level at specified frequencies. 1.1 Input power measurement (amplifiers only). 1.2 Test circuit. FIGURE 4250-1. Input power measurement. 1.3 Procedure. With the TUT inserted in the circuit shown on figure 4250-1, part of the power from the signal source is fed by directional coupler to a power measuring device. The power (Pm) is then measured by either test A, B, or C as appropriate. The input power to the TUT is then determined from the following relationship: Pi = Pm + 10 log [(antilog k/10) -1] + Lf Where:

Pi = input power in dBm to the TUT (the power level to be determined). Pm = power level actually measured in dBm. k = coupling coefficient of the directional coupler in dB. Lf = attenuation in dB of filter at measurement frequency. NOTE: If the coupling coefficient k is 20 dB or higher, the following simplified expression may be used with negligible error: Pi = Pm + k 2. Output power measurement. 2.1 Test circuit. FIGURE 4250-2. Output power measurement. METHOD 4250A 4250-1 25 March 1970