Downloaded from https://www.everyspec.com MIL-STD-1311C 31 January 2001 2.2 Procedure. With the TUT inserted in the circuit shown on figure 4250-2, the output of the TUT is dissipated in a broadband load. A portion of the TUT output is fed by directional coupler to an attenuator, then to a power measuring device, and then measured by either test A, B, or C as appropriate. The combination of coupling and attenuation shall be such that the power measuring device is operating in its proper range. The output power of the TUT is then determined from the following relationship: Po = k + A + Pm + Lf Where: Po = output power in dBm. k = coupling coefficient of directional coupler in dB.

A = attenuation in dB of attenuator. Pm = power level actually measured in dBm. Lf = attenuation in dB of filter at measurement frequency. 3. Test A (calorimeter method of average power measurement). 3.1 Test circuit. FIGURE 4250-3. Calorimeter method. 3.2 Procedure. Using the circuit shown on figure 4250-3, the power to be measured is dissipated in the calorimeter. The power is then determined from the following relationship: (1) Po = 4.187 x m x Sp x T Where: Po = dissipated power in watts. m = coolant flow in grams per second. Sp = specific heat of coolant in calories per gram per C. T = temperature rise of coolant in C. METHOD 4250A 25 March 1970 4250-2