US Army Developmental Test Command TOP 1-2-622 — Test Setup #1

Downloaded from https://www.everyspec.com TOP 1-2-622 11 September 2009 APPENDIX E: Overview of VEMP test instrumentation and mathematics Provided is an overview of the data processing that occurs in VEMP testing to form a basic knowledge foundation. It should be noted that this is not an in-depth presentation and that the procedures used and data accuracy provided by the instrumentation are in general the responsibility of the test facility. 1. DATA MEASUREMENTS: For testing at pulse/high frequency facilities most sensors and data links are self calibrated by the instrumentation specialist.

The traceability back to a laboratory calibration is through the use of a network analyzer which has calibration performed at the factory or a standard calibration laboratory. The instrumentation is basically divided into three sections which will be discussed separately below: SENSORS: In general the facility calibration of sensors is limited to the ones which measure voltage or current. Sensors for measuring fields (electric or magnetic) or for measuring surface currents are factory calibrated. The current probes are available with different current measuring capabilities and different through hole sizes; inmost cases (except for very small very specific probes) clamp on devices are used.

A special adapter is required for each size probe for the facility calibration to be performed. The probe adapter (see Figure E-1) allows for the placement of a probe around the center conductor and inside the external conductor. The probe is then connected to a network analyzer. The network analyzer output feeds both the forcing function and the reference measurement and measures the voltage output of the current probe. This forcing function is driven across the frequency spectrum for which the current probe is being calibrated.

The current probe voltage output is divided by the current through the fixture; the current through the fixture is based on the measured reference voltage and the fixed load impedance. The result is the transfer function of the current probe versus frequency which is called the probe transfer impedance since the units are volts/ampere. The voltage probes are calibrated in the same manner with the exception that the transfer function units are volts/volt. Experience has indicated that 400 measurements across the frequency spectrum tend to be adequate for calibration. This frequency calibration is used as follows:

The data produced by the current or voltage probe is recorded on fast oscilloscopes as voltages in the time domain. The recorded data then requires correction for the probe transfer function (frequency response) and conversion to the correct units of measure. This is accomplished using convolution in the time domain, the process however takes place in the frequency domain since in the frequency domain convolution reduces to simple point by point division. The process for a current probe is that the recorded time domain waveform is converted to the frequency domain using Fast Fourier Transforms (FFT).

The resulting frequency data is then divided on a per frequency basis by the probe transfer function at that frequency, interpolation between probe calibration frequency points is used to match the exact frequency resulting from the FFT. Once the convolution has taken place the FFT is now corrected for probe response and the inverse FFT results in the corrected time response with the appropriate units E-1