Тексел ТХ01-RS — 6.3.1 Tachometer mode

2 : the pulse duration is less than minImp – the signal is not passed by the bounce protection system; 3 : the pulse duration is greater than minImp , but the pause duration is shorter than minImp – the signal is not skipped by the chatter protection system. Figure 6.4 – Parameter «minImp» 6.3.1 Tachometer mode When performing the function of a tachometer: • device measures the instantaneous value of the duration of the time intervals (T) between the leading or trailing edges (in s); • device calculates the value of N / T, where N is the value determined by the dimension of the display of the measured value; • device displays the obtained value taking into account the scaling factor. 33