An eighth-order moving average filter is used to damp the signal in the device. The MAV.L parameter defines the minimum time (in seconds) for filling the filter buffer. Adding a new value to the buffer will be done not earlier than in ( MAV.L / 8) seconds. At input frequencies less than or equal to (8 / MAV.L ), the settling time of the input signal to the digital signal can be longer than MAV.L . By default, this filter is enabled. 6.3.2 Operating time counter The device counts the operating time if the signal level on the operating time counter reaches a logical 1. When the device performs this function, the To see the current operating time on display, press and hold the button .
If you release this button, the display returns to the tachometer reading. The measured operating time is shown on the display (see Table 6.1 for details). If the current range is exceeded, the device automatically sets the next display range of the measurement results. The maximum value of the operating time takes values from 1 s to 9999 days 23 hours. For devices with discrete output, it is possible to set the limit values of the operating time, upon reaching which the output is switched on. LED lights up. 35